Hanna Instruments HI 83200C User Manual
Page 60

118
119
Silver
• Press the
Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Fill a second cuvette up to the mark with 10 mL of the reacted sample.
• Add 3 drops of Dispersing Agent (HI 93703-51),
replace the cap and invert gently to mix for about 10
seconds.
• Insert the second cuvette into the instrument.
• Press
Read to start the reading. The instrument displays the results in mg/L of silver.
INTERFERENCES
Interference may be caused by:
Al
3+
above 30 mg/L
Fe
2+
above 1.5 mg/L
Ca
2+
above 1000 mg/L as CaCO
3
Fe
3+
above 10 mg/L
Cd
2+
above 20 mg/L
K
+
above 500 mg/L
Cl¯ above 8000 mg/L
Mn
2+
above 25 mg/L
Co
2+
above 1.5 mg/L
Mg
2+
above 1000 mg/L as CaCO
3
Cr
3+
above 20 mg/L
Na
+
above 5000 mg/L
Cr
6+
above 40 mg/L
Ni
2+
above 1.5 mg/L
Cu
2+
above 15 mg/L
Pb
2+
above 20 mg/L
F¯ above 20 mg/L
Zn
2+
above 30 mg/L
Silver
• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second
beaker (the sample) and swirl gently to mix. Press
Timer and the
display will show the countdown prior to adding reagent C or,
alternatively, wait for 2 minutes.
• Then add exactly 1.0 mL of HI 93737C-0 Indicator reagent to
each beaker and swirl. Press
Timer or, alternatively, wait for 2
minutes.
• Then, in both cases, add 1.0 mL of HI 93737D-0 Fixing reagent
to each beaker and swirl. Press
Timer or, alternatively, wait for
2 minutes.
• Fill one cuvette up to the mark with 10 mL of the blank.
• Add 3 drops of Dispersing Agent (HI 93703-51), replace the cap
and invert gently to mix for about 10 seconds.
• Place the cuvette into the holder and close the lid.
# 2
# 2
# 1
# 2
# 1
# 1
# 1
10 mL
10 mL
# 2
# 2