70

71

Iron LR

IRON LOW RANGE

SPECIFICATIONS

Range

0 to 400

µ

g/L

Resolution

1

µ

g/L

Accuracy

±10

µ

g/L ±8% of reading at 25 °C

Typical EMC

±1

µ

g/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the TPTZ Method. The reaction between iron and the reagent causes a
violet tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93746-0

TPTZ Reagent

2 packets

REAGENT SETS

HI 93746-01 Reagents for 50 tests
HI 93746-03 Reagents for 150 tests
For other accessories see page 128.

MEASUREMENT PROCEDURE

• Select the

Iron LR method using the procedure described in

the

Method Selection section (see page 12).

• Fill one graduated mixing cylinder up to the 25 mL mark

with deionized water.

• Add the content of one packet of HI 93746-0 TPTZ

reagent, close the cylinder and shake vigorously for 30
seconds. This is the blank.

• Fill a cuvette with 10 mL of the blank up to the mark and

replace the cap.

• Place the cuvette into the holder and close the lid.

Iron HR

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 3 minutes and press

Read. When the timer ends the meter will perform the reading.

• The instrument displays concentration in

mg/L of iron.

INTERFERENCES

Interference may be caused by:
Molybdate Molybdenum above 50 ppm
Calcium above 10000 ppm (as CaCO

3

)

Magnesium above 100000 ppm (as CaCO

3

)

Chloride above 185000 ppm.

12

3

6

9

30"

10 mL

25 mL