Hanna Instruments HI 83200C User Manual
Page 36

70
71
Iron LR
IRON LOW RANGE
SPECIFICATIONS
Range
0 to 400
µ
g/L
Resolution
1
µ
g/L
Accuracy
±10
µ
g/L ±8% of reading at 25 °C
Typical EMC
±1
µ
g/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 575 nm
Method
Adaptation of the TPTZ Method. The reaction between iron and the reagent causes a
violet tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93746-0
TPTZ Reagent
2 packets
REAGENT SETS
HI 93746-01 Reagents for 50 tests
HI 93746-03 Reagents for 150 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
Iron LR method using the procedure described in
the
Method Selection section (see page 12).
• Fill one graduated mixing cylinder up to the 25 mL mark
with deionized water.
• Add the content of one packet of HI 93746-0 TPTZ
reagent, close the cylinder and shake vigorously for 30
seconds. This is the blank.
• Fill a cuvette with 10 mL of the blank up to the mark and
replace the cap.
• Place the cuvette into the holder and close the lid.
Iron HR
• Reinsert the cuvette into the instrument.
• Press
Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 3 minutes and press
Read. When the timer ends the meter will perform the reading.
• The instrument displays concentration in
mg/L of iron.
INTERFERENCES
Interference may be caused by:
Molybdate Molybdenum above 50 ppm
Calcium above 10000 ppm (as CaCO
3
)
Magnesium above 100000 ppm (as CaCO
3
)
Chloride above 185000 ppm.
12
3
6
9
30"
10 mL
25 mL