Hanna Instruments HI 83200C User Manual
Page 42

82
83
Molybdenum
• Press
Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for five minutes and press
Read. When the timer ends the meter will perform the reading. The
instrument displays concentration in
mg/L of molybdenum.
• Press ▲ or ▼ to access the second level functions.
• Press the
Chem Frm key to convert the result in mg/L of molybdate (MoO
4
) and sodium molybdate
(Na
2
MoO
4
).
• Press ▲ or ▼ to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Aluminum above 50 mg/L
Chromium above 1000 mg/L
Copper above 10 mg/L
Iron above 50 mg/L
Nickel above 50 mg/L
Nitrite, as NO
2
¯
Sulfate above 200 mg/L
Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents.
• Fill one graduated mixing cylinder up to the 25
mL mark with the sample.
• Add the content of one packet of HI 93730A-0
molybdenum reagent, close the cylinder and invert
it several times until completely dissolved.
• Add the content of one packet of HI 93730B-0
molybdenum reagent to the cylinder, close and
invert it several times until completely dissolved.
• Add the content of one packet of HI 93730C-0
molybdenum reagent to the cylinder, close and
shake it vigorously.
• Fill an empty cuvette with 10 mL of sample up to
the mark and replace the cap.
• Insert the cuvette into the instrument.
Molybdenum
25 mL
10 mL