Hanna Instruments HI 83208-2008 User Manual
Page 34
66
67
Silver
• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second beaker
(the sample) and swirl gently to mix.
• Press TIMER and the display will show the countdown prior to adding reagent C or, alternatively, wait
for 2 minutes. Then add exactly 1.0 mL of HI 93737C-0 Indicator reagent to each beaker and swirl.
• Press TIMER or, alternatively, wait for 2 minutes. Then add 1.0 mL of HI 93737D-0 Fixing reagent to
each beaker and swirl.
• Then press TIMER or, alternatively, wait for 2 minutes to allow reaction to complete.
• Fill one cuvette up to the mark with 10 mL of the blank.
• Place the cuvette into the holder and close the lid.
# 2
# 2
# 1
# 2
10 mL
# 1
# 1
# 1
10 mL
Silver
SILVER
SPECIFICATIONS
Range
0.000 to 1.000 mg/L
Resolution
0.001 mg/L
Accuracy
±0.005 mg/L ±10% of reading
Typical EMC
± 0.001 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 575 nm.
Method
Adaptation of the PAN method. The reaction between silver and reagents causes an
orange tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93737A-0
Buffer Reagent A
1 mL
HI 93737B-0
Buffer Reagent B
1 mL
HI 93737C-0
Indicator Reagent C 2 mL
HI 93737D-0
Fixing Reagent D
2 mL
HI 93703-51
Dispersing Agent
4-6 drops
(only when necessary, see note)
REAGENT SETS
HI 93737-01 Reagents for 50 tests
HI 93737-03 Reagents for 150 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
• Select the
Silver method using the procedure described in the
Method Selection section (see page 12).
Note: for best results perform your tests between 20-24°C.
• Fill two graduated beakers with 25 mL of sample.
• Add 1.0 mL of HI 93737A-0 Buffer reagent to one beaker (the
blank) and swirl gently to mix.
# 1
# 2
# 1