beautypg.com

Hanna Instruments HI 83208-2008 User Manual

Page 21

background image

40

41

Manganese LR

• Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagent

solution to each cuvette, replace the caps and shake gently.

• Add 1 mL of the HI 93748C-0 0.1% PAN indicator solution

to each cuvette, replace the caps and shake gently.

• Place the cuvette with the reacted deionized water

(blank) into the holder and close the lid.

• Press TIMER and the display will show the countdown prior

to zeroing. Alternatively wait for 2 minutes and then press
ZERO. The display will show "-0.0-" when the meter is
zeroed and ready for measurement.

• Insert the second cuvette with the reacted sample into

the instrument.

Manganese LR

MANGANESE LOW RANGE

SPECIFICATIONS

Range

0 to 300 µg/L

Resolution

1 µg/L

Accuracy

±2 µg/L ±3% of reading

Typical EMC

±1 µg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the PAN Method. The reaction between manganese and the reagents
causes an orange tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93748A-0

Ascorbic acid

2 packets

HI 93748B-0

Alkaline-cyanide sol.

0.40 mL

HI 93748C-0

0.1% PAN indicator

2 mL

HI 93703-51

Dispersing Agent

4-6 drops (only when necessary, see note)

REAGENT SETS

HI 93748-01 Reagents for 50 tests
HI 93748-03 Reagents for 150 tests
For other accessories see page 73.

MEASUREMENT PROCEDURE

• Select the

Manganese LR method using the procedure described in

the

Method Selection section (see page 12).

• Fill one cuvette with 10 mL of deionized water (up to the mark).

• Fill a second cuvette with 10 mL of sample (up to the mark).

• Add the content of one packet of HI 93748A-0 Ascorbic acid to

each cuvette, replace the caps and shake gently until completely
dissolved.

10 mL

10 mL