Hanna Instruments HI 83208-2008 User Manual
Page 16
30
31
Fluoride
FLUORIDE
SPECIFICATIONS
Range
0.00 to 2.00 mg/L
Resolution
0.01 mg/L
Accuracy
±5% of reading
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 575 nm
Method
Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition, SPADNS method. The reaction between fluoride and the liquid reagent
causes a red tint in the sample.
REQUIRED REAGENT
Code
Description
Quantity
HI 93729-0
SPADNS Reagent
4 mL
REAGENT SETS
HI 93729-01 Reagents for 100 tests
HI 93729-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
• Select the
Fluoride method using the procedure described in
the
Method Selection section (see page 12).
• Add 2 mL of HI 93729-0 SPADNS reagent to two cuvettes.
• Fill one of the cuvettes with distilled water up to the mark, replace
the cap and invert several times to mix.
• Fill the other cuvette with sample up to the mark, replace the cap
and invert several times to mix.
• Place the cuvette with the reacted distilled water (# 1) into the
holder and close the lid.
10 ml
10 ml
# 1
# 2
2 ml
# 1
Copper LR
• Remove the cuvette.
• Add the content of one packet of HI 93747-0 Bicinchoninate.
Replace the cap and shake gently for about 15 seconds.
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to
the measurement or, alternatively, wait for 45 seconds and press
READ. When the timer ends the meter will perform the reading.
• The instrument displays the results in mg/L of copper.
INTERFERENCES
Interference may be caused by:
Silver
Cyanide
For samples overcoming buffering capacity of reagent (around pH 6.8), pH should be adjusted between 6
and 8.