26

27

Copper HR

COPPER HIGH RANGE

SPECIFICATIONS

Range

0.00 to 5.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.02 mg/L ±4% of reading

Typical EMC

±0.01 mg/L

Deviation

Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the

EPA method. The reaction between copper and the bicinchoninate

reagent causes a purple tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93702-0

Bicinchoninate

1 packet

REAGENT SETS

HI 93702-01 Reagents for 100 tests
HI 93702-03 Reagents for 300 tests
For other accessories see page 73.

MEASUREMENT PROCEDURE

• Select the

Copper HR method using the procedure described in

the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to the mark)

and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed

and ready for measurement.

10 ml

• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait

for 2 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading.

• The instrument displays the results in mg/L of total chlorine.

Note: free and total chlorine have to be measured separately with fresh unreacted samples following the
related procedure if both values are requested.

INTERFERENCES

Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample for approximately 2

minutes after adding the powder reagent.
In case of water with alkalinity greater than 250 mg/L CaCO

3

or acidity greater than 150 mg/L CaCO

3

,

the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.