Nickel HR

• Reinsert the cuvette into the instrument.

• Press TIMER and the display will show the countdown prior

to the measurement or, alternatively, wait for 1 minute and
press READ. When the timer ends the meter will perform the
reading.

• The instrument displays concentration in g/L of nickel.

INTERFERENCES

Interference may be caused by copper.

Nickel HR

SPECIFICATIONS

Range

0.00 to 7.00 g/L

Resolution

0.01 g/L

Accuracy

±4% of reading

Typical EMC

±0.02 g/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the photometric method. The reaction between nickel and the reagent
causes a blue tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93726-0

Powder reagent 1 packet

REAGENT SETS

HI 93726-01 Reagents for 100 tests
HI 93726-03 Reagents for 300 tests
For other accessories see page 73.

MEASUREMENT PROCEDURE

• Select the

Nickel HR method using the procedure described in

the

Method Selection section (see page 12).

• Fill the cuvette up to the mark with 10 mL of unreacted

sample and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press ZERO key. The display will show “-0.0-” when the meter is

zeroed and ready for measurement.

• Remove the cuvette and add the content of one packet of HI 93726-0

reagent. Replace the cap and shake gently until completely dissolved.

10 mL

NICKEL HIGH RANGE