Hanna Instruments HI 2216 User Manual
Page 39

39
LODPxxx Requests the xxx
th
pH record logged data.
LODMxxx Requests the xxx
th
mV/Rel mV record logged data.
LODIxxx
Requests the xxx
th
ISE record logged data.
LODPALLff Requests all pH Log on demand - ff=frame number
LODMALLff Requests all mV/Rel mV Log on demand - ff=frame number
LODIALLff Requests all ISE Log on demand - ff=frame number
Note: The log on demand can be downloaded in frames of 8
records (01 is first frame number)
LLSxff: Requests information about all lots in frames of 10 lots (a frame
contains information about 10 lots)
Command Parameters:
x - range (see Note)
ff - requested frame number - first frame is labeled 01
GLDxxxff: Requests the records of the “xxx” lot number. The records are
sent in frames of 10 records; “ff” is the frame number (01 first
frame). (Example: Lot 13 has 53 records. The records will be
sent in 6 frame, 5 with10 records and 1 with 3 records.)
Command Parameters:
xxx - Lot number (eq: for lot number 1 xxx = 001)
ff - requested frame number - first frame is labeled 01
Errors:• “Err3” is sent if the Log on demand is empty.
• “Err4” is sent if the requested set parameter is not available.
• “Err5” is sent if an argument of the command is not correct.
• “Err6” is sent if the requested range is not available.
• “Err7” is sent if the instrument is in logging mode.
• “Err8” is sent if the instrument is not in measurement mode.
• Invalid commands will be ignored.
Note: P - request for pH range.
M - request for mV and Rel mV ranges.
I - request for ISE range.