PC INTERFACE

PC INTERFACE

PC INTERFACE

PC INTERFACE

PC INTERFACE

Data transmission from the instrument to the PC can be done with the
HI 92000 Windows

®

compatible software (optional).

HI 92000 also

offers graphing and an on-line help feature.
Data can be exported to the most popular spreadsheet programs for
further analysis.
To connect your instrument to a PC, use a standard USB cable connector.
Make sure that your instrument is switched off and plug one connector to
the instrument’s USB socket and the other to the USB port of your PC.

Note: If you are not using Hanna Instruments HI 92000 software,

please see the following instructions.

SENDING COMMANDS FROM PC

It is also possible to remotely control the instrument with any terminal
program. Use a standard USB cable to connect the instrument to a PC,
start the terminal program and set the communication options as follows:
8, N, 1, no flow control, 9600 baud rate.

COMMAND TYPES

To send a command to the instrument follow the next scheme:

>

where:

is a 16 ASCII character
is the command code.

Note: Either small or capital letters can be used.

SIMPLE COMMANDS

RNG

Is equivalent to pressing

RANGE

CAL

Is equivalent to pressing

CAL

CFM

Is equivalent to pressing

CFM

UPC

Is equivalent to pressing the

UP arrow key

DWC

Is equivalent to pressing the

DOWN arrow key

LOG

Is equivalent to pressing

LOG

RCL

Is equivalent to pressing

RCL

SET

Is equivalent to pressing

SETUP

Notes:• If the reading is not close to the selected calibration point,

WRONG” tag will blink. Verify calibration condition or

contact your vendor if you can not calibrate.

• Pressing

CAL key during temperature or mV calibration the

instrument quit calibration mode and restart to measurement
mode without memorizing calibration.