L O G G I N G F U N C T I O N

L O G G I N G F U N C T I O N

L O G G I N G F U N C T I O N

L O G G I N G F U N C T I O N

L O G G I N G F U N C T I O N

Up to 700 LOG samples can be stored into memory.
200 manually logged records and 500 logging records can be stored in
the memory. To select logging type enter SETUP menu.

LOGGING THE CURRENT DATA

(manual logging)

Select the manual logging mode in SETUP menu.
To store the current reading into memory press

LOG while the instrument is in

measurement mode.
The instrument will display “

MAn” on the primary LCD, the record

number on the secondary LCD and “

LOG” tag for a few seconds (see

example below: record No. 5):

followed by the number of free records:

If there are less than 6 memory locations remaining, the record number
and “

Lo” message will be displayed to alert the user.

If the log space is full (200 records), “

FULL LOG” message will be

displayed and no more data will be saved.

When

LOG is pressed, a complete set of information is stored: date, time,

pH, mV, temperature and calibration data.