Page 24

background image

24

LOT LOGGING

Select “

StAb” (stability logging) or the desired time interval.

To start interval logging press

LOG key while the instrument is in

measurement mode.
When the selected interval is reached, the instrument will display the
current lot number on the primary LCD line, the record number on the
secondary LCD line and the “

LOG” tag (see example below: Lot 5 record 7)

followed by the number of free records on the corresponding memory space.

If stability logging is selected, a complete set of data is memorized every
time the reading becomes stable after an unstable condition.
To stop interval logging press

LOG key again.

Note: When pressing any key that is not active, while Log interval is

running, the following message is displayed for a few seconds:

VIEW LOGGED DATA

Press

RCL while in measurement mode to retrive the stored information.

If no data were logged for the current selected measurement range and
no lots are memorized, one of the next messages will be displayed:

No pH measurement records:

No Relative mV and mV records: