• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected

buffer, the “

READY” tag will be displayed and the

CFM” tag will blink.

• Press

CFM to confirm calibration.

• The calibrated value is then displayed on the

primary LCD and the third expected buffer value on
the secondary LCD.

• After the second calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”) into
the next buffer solution and stir gently. The temperature probe
should be close to the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, the

READY” tag will be displayed and the “CFM” tag

will blink.

• Press

CFM to confirm calibration.

• The calibrated value is then displayed on the

primary LCD and the fourth expected buffer value
on the secondary LCD.

• After the third calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”) into
the next buffer solution and stir gently. The temperature probe should
be close to the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, the

READY” tag will be displayed and the “CFM” tag

will blink.

• Press

CFM to confirm calibration.

• The calibrated value is then displayed on the

primary LCD and the fifth expected buffer value on
the secondary LCD.

• After the fourth calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”) into
the next buffer solution and stir gently. The temperature probe
should be close to the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.