• Meter status (2 chars of status byte): represents a 8

bit hexadecimal encoding.
• 0x10 - temperature probe is connected
• 0x01 - new GLP data available
• 0x02 - new SETUP parameter

• Reading status (2 chars): R - in range, O - over

range, U - under range. First character corresponds
to the appropriate range reading. Second character
corresponds to mV reading.

• Primary reading (corresponding to the selected range)

- 7 ASCII chars, including sign and decimal point.

• Secondary reading (only when primary reading is not

mV) - 7 ASCII chars, including sign and decimal point.

• Temperature reading - 8 ASCII chars, with sign and

two decimal points, always in ºC.

MDR

Requests the instrument model name and firmware code
(16 ASCII chars).

GLP

Requests the calibration data record.

The answer string contains:

• GLP status (1 char): represents a 4 bit hexadecimal

encoding.
• 0x01 - pH calibration available
• 0x02 - Rel mV calibration available
• 0x04 - ISE calibration available

• pH calibration data (if available), which contains:

• the number of calibrated buffers (1 char)
• the ion charge, with sign (2 chars)
• the offset, with sign and decimal point (7 chars)
• the average of slopes, with sign and decimal

point (7 chars)

• the calibration time,

yymmddhhmmss (12 chars)

• buffers information (for each buffer)

• type (1 char): 0 - standard, 1 - custom
• status (1 char): N (new) - calibrated in last

calibration; O (old) - from an old calibration.

• warnings during calibration (2 chars): 00 -

no warning, 04 - Clean Electrode warning.

• buffer value, with sign and decimal point (7

chars).