Default slope value (mV/decade):
–59.16 (monovalent cation) - ion charge is 1

59.16 (monovalent anion) - ion charge is –1

–29.58 (divalent cation) - ion charge is 2

29.58 (divalent anion) - ion charge is –2
100 - ion charge is “

undF”

• Press

CLR during calibration if you want to clear calibration

values. The instrument will display “

CLR” message and will

return to measurement mode.

• The calibration needs to be cleared when a new electrode is

attached to the meter.

• The instrument will display “

----” on the primary LCD if it is

not calibrated or after all calibrations are cleared.

• Press

RANGE to display the temperature reading on

the LCD during calibration.

ONE-POINT CALIBRATION

• Press

CAL after first calibration point was confirmed. The instrument

memorizes the one-point calibration information and returns to
measurement mode.

• Press

RANGE to display the temperature reading on

the LCD during calibration.