Hanna Instruments HI 2216 User Manual
Page 18

18
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
GOOD LABORATORY PRACTICE (GLP)
RELATIVE
RELATIVE
RELATIVE
RELATIVE
RELATIVE m
m
m
m
m V CALIBRATION
V CALIBRATION
V CALIBRATION
V CALIBRATION
V CALIBRATION
GLP is a set of functions that allows storage and retrieval of data
regarding the maintenance and status of the electrode.
All data regarding pH, Rel mV or ISE calibration is stored for the user to
review when necessary.
EXPIRED CALIBRATION
This instrument allows the user to set the number of days before the next
required calibration. This value can be set from 1 to 7 days. The default
setting is off (disabled). The “
CAL” “INTV” tags will start blinking to
warn the user that the instrument should be recalibrated.
For example, if a 4 days time out has been selected, the instrument will
issue the alarm exactly 4 days after the last calibration.
If the expiration value is changed (e.g. to 5 days), then the alarm will
be immediately recalculated and appear 5 days after the last
calibration.
• Press
CAL when the instrument is in RELATIVE mV measurement
mode. The “
mV” and “ ” tags will be displayed. Absolute mV is
displayed on the primary LCD and “
AbS” message is displayed on the
secondary LCD.
• When the absolute reading is stable and in measurement range, the
instrument asks for confirmation.
• If the reading is out of range, “
WRONG” tag will be displayed.
• Press
CFM to confirm the absolute value. The instrument will display
“
0.0 mV“ on the primary LCD and “rEL” message on the secondary
LCD. In this moment the relative mV offset is equal to absolute mV
reading.
• Use the
ARROW keys if you want to change the displayed relative
mV value.
• Press
CFM to confirm the relative mV value. The relative mV
offset is displayed on the primary LCD. The instrument returns to
measurement mode.
Note: The relative mV value can be changed only inside the relative mV
offset window (± 2000 mV).