Turning off the unit, Turning on the unit – EXFO QA-805 User Manual

Page 32

background image

Getting Started

24

QA-805 and W2CM Interface Module

Turning the Unit On/Off

Turning Off the Unit

To safely turn off the QA-805, please ensure the following steps

are taken:
1.
Stop all test applications and save results to prevent measurement loss.

2. Power down the Processor v1(AM4020M System Controller) by pulling

out the HotSwap switch to its first click position. Wait until the Blue
HotSwap LED stops flashing and turns Solid Blue to indicate that the
System Controller has shut down safely.

3. Power down the Hard Drive(AM4500) by pulling out the HotSwap

switch. Wait until the Blue HotSwap LED stops flashing and turns Solid
Blue to indicate that the Hard Drive has shut down safely.

4. Power down all active power supplies.

Turning On the Unit

To safely turn on the QA-805, please ensure the following steps

are taken:
1.
Ensure that all empty slots are filled with the appropriate sized Filler

Card/Air Baffle.

2. Ensure that the HotSwap switches on both the Processor v1(AM4020M

System Controller) and Hard Drive(AM4500) are fully pushed in.

3. Power up all required power supplies.