Installation Instruction Warnings

Installation Instruction Warnings

Rack Mounting

If rack mounted units are installed in a closed or multi-unit rack assembly,
they may require further evaluation by Certification Agencies. The
following items must be considered:

1. The ambient temperature within the rack may be greater than room

temperature. Installation should be such that the amount of air flow
required for safe operation is not compromised. The maximum
temperature for the equipment in this environment is 40 ºC.
Consideration should be given to the maximum rated ambient.

2. Installation should be such that a hazardous stability condition is not

achieved due to uneven loading.

3. Reliable earthing of rack-mounted equipment should be maintained.

4. Thumbscrews should be tightened with a tool after both initial

installation and subsequent access.

Input Supply

Check nameplate ratings to assure there is no overloading of supply
circuits that could have an effect on overcurrent protection and supply
wiring.

Unit is provided with multiple power inputs. To disconnect all from the
unit remove all incoming power sources.

C

AUTION

Do not lift the chassis by the fan tray handle.