Measuring Power or Loss (EPM-50)

10

Measuring Power or Loss

Measuring absolute power is done the same way as referencing the power meter to
a source, except for the referencing step.

To perform power or loss measurements:
1.
If necessary, perform an offset nulling (see Nulling Electrical Offsets on page 8).
2. Check and clean your fibers appropriately for optimum performance

(see Cleaning and Connecting Optical Fibers on page 7).

3. For loss measurements, reference your power meter to a light source

(see Referencing Your Power Meter to a Source on page 9), then deactivate the
light source.

4. Using the proper adapter and test jumpers, connect a fiber under test to a light

source (such as ELS-50) and to the detector port of your unit.

Note:

If you have referenced your power meter to a source, simply connect a
fiber under test to the test jumpers used for referencing.

Fiber under test

Test

jumper

Test

jumper

Light

source

Power

meter

Adapter

Bulkhead

connector

Bulkhead