beautypg.com

3 analogue output, Figure 7: eex d conduit installation, Enmet corporation – ENMET EX-6165 User Manual

Page 8

background image

EX-6120, 6165 SERIES IR

ENMET Corporation

5

Figure 5: Unplugging the Display Board for Connection Purposes

3.2 Power Supply

The

EX-6120, 6165 S

ERIES

IR

is powered from an 8 – 24 volt dc, 5 W maximum output supply. This may be either:-

a)

An independent supply powering one or more units.

b)

A supply that is an integral part of a control card within a monitoring panel.

3.3 Analogue Output

The analogue output provides a means of indicating to external equipment (e.g. data loggers, remote displays, control
cards) the gas levels currently being detected by the system. The output is in the form of the industry standard 4 to
20mA current source.

The analogue output is factory set such that 4mA represents zero gas and 20mA represents the full scale gas level of the

EX-6120, 6165 S

ERIES

IR

.

Figure 6: Analogue Output

3.4 RS232 Output

This output is designed to be compatible with a PC fitted with an RS232 serial communications port. Suitable software
is available for installation in the PC. Contact

ENMET

Corporation for details.

Figure 7: EEx d Conduit Installation

Important

Connect enclosure

earth-ground to a high

integrity earth-ground

ATEX Certified compound-filled

conduit seals (stopping boxes)

- +

PSU

- +

Aout

0

V

R

x

T

x

RS232

Conduit

Conduit

Mount vertically

as shown

N

/O

C

O

M

N

/C

Relay 1
Alarm 1

N

/O

C

O

M

N

/C

Relay 2
Alarm 2

N

/O

C

O

M

N

/C

Relay 3

Fault

EX-6120,

6165

0V

A+

0V
12-24V DC

Control

Unit

I (4-20mA)

PSU

This manual is related to the following products: