EX-6120, 6165 SERIES IR

ENMET Corporation

18

11.2.16 E : 20 – Negative Zero Suppression

This option is used to allow the user to suppress small amounts of negative sensor zero drift. The setting can be set between 0 and
10% of the sensor range as set by the FSD value.

Press MENU to open the menu system.

Using the NEXT and PREVIOUS buttons, select menu option: E:20

Press ENTER.

Using the INCREASE and DECREASE buttons, set the required zero suppression value.

Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button exits without any change.

Press MENU to close the menu system.

11.2.17 E : 27 – Hysteresis

This option sets the alarm hysteresis. The relays will chatter if this level is set to 100.

95% is the usual setting for pellistor sensors.

11.2.18 E : 28 – Firmware update

Do Not Enter this Menu

11.2.19 E : 29 – Relay 3

The unit is fitted with a relay that is operated in conjunction with a fault condition. The user can select if the relay is
normally Energized, ‘E’ or normally de-energized, ‘d’ when the unit is not in a fault condition.

From the menu system select menu option: E: 29 and press ENTER.

The display will show the following:

E:r

Normally energized, rising alarm

d:r

Normally de-energized, rising alarm

E:F

Normally energized, falling alarm

d:F

Normally de-energized, falling alarm

The mode of operation can be changed by pressing the UP button.

Press ENTER to accept the new relay mode of operation.
N

OTE

: Pressing MENU instead of ENTER leaves the unit without change.

Press MENU to return the instrument to its standard mode of operation.

11.2.20 E : 30 – Password

Use this menu option to enable / disable the password feature.

Place the

EX-6120, 6165 S

ERIES

IR

in the password menu as follows:

Press the MENU to open the menu system.

Using the NEXT and PREVIOUS buttons, select menu option: E:30

Press ENTER.

The display shows either On or OFF

Press the UP button to select the desired setting.

Press ENTER to accept the setting.
N

OTE

: Pressing MENU instead of ENTER leaves the unit without change.

Press MENU to return the instrument to its standard mode of operation.

Note: If the password is in operation then the user will be prompted with PASS when ever the menu key is pressed. Pressing

the MENU key again will result in the restricted user access, i.e. only the zero and span options will be available.
Entering the correct password will give access to the full menu facility.

11.2.21 E : 31 – Warm-up timer

The sensor goes through a stabilization period when power is first applied. This option allows the user to adjust the time
that is applied before readings are displayed / transmitted.

From the menu system select menu option: E: 31 and press ENTER.

Using the INCREASE and DECREASE buttons, set the required warm-up time.

Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button exits without any change.

Normally warm time is 45 seconds.