EX-6120, 6165 SERIES IR

ENMET Corporation

19

Appendix A

A.1 Analog Output Calibration

The analogue output is that of a current source. In order to calibrate the output it is necessary to monitor the output
signal. This can be performed in one of two ways:

Current measurement:

Connect an ammeter (or multimeter set to measure current in the mA range) in series with the analogue output.

Advantage:

Direct measurement of output.

Disadvantage: The analogue output has to be disconnected to allow the connection of the meter.

Voltage measurement:

Permanently connect a 10R resistor in series with the analogue output. When calibration is required, connect a
voltmeter
(or multimeter set to measure voltage in the mV range) across the 10R resistor.

Advantage:

No need to disturb wiring between

EX-6120, 6165 S

ERIES

IR

and control unit.

Disadvantage: Measurement accuracy dependent upon resistor tolerance.

A.1.1 Analogue Output Zero

Monitor the current sourced from the analogue output of the EX-6120, 6165 S

ERIES

IR using a suitable method.

From the menu system select menu option: E: 4 and press ENTER.

Using the INCREASE and DECREASE buttons; adjust the output to the required level (4mA or 40mV).

Press ENTER.
N

OTE

: Pressing the MENU button rather than the ENTER button exits the zero feature without performing the

calibration.

Press the MENU button to close the menu system.
N

OTE

: The ZERO factor will be displayed momentarily on exit.

A.1.2 Analogue Output Span

Always zero the analogue output prior to performing a span operation.

From the menu system select menu option: E: 5 and press ENTER.

Using the INCREASE and DECREASE buttons; adjust the output to the required level.

Press ENTER button.
N

OTE

: Pressing the MENU button rather than the ENTER button exits the span feature without performing the calibration.

Press the MENU button to close the menu system.
N

OTE

: The ANALOGUE OUTPUT SPAN factor will be displayed momentarily on exit.


A.2 Error Codes

Not applicable to the Pellistor Catalytic Sensor variant


+

FGD10
Analogue Output
(4 – 20mA)

Control

Unit

mA

+

Meter range:
0.004A – 0.020A
4mA – 20mA


+

FGD10
Analogue Output
(4 – 20mA)

Control

Unit

10R

mV

+

Meter range:
0.04V – 0.20V
40mV – 200mV