EX-6120, 6165 SERIES IR

ENMET Corporation

9

5.0 Calibration

5.1 Sensor Calibration

This feature allows the sensor to be calibrated. Ensure that the correct sensor type is selected in the configuration prior to
calibration. Refer to section for details of the menu system operation.

5.1.1 Sensor Zero

Tap the magnet over the

menu

switch, the display will change to Pass, Tap the magnet again on

menu

display will

show E 1.

Tap the magnet over the

enter

switch the display will then alternate E: 01 Wait approximately 1 minute and Tap

enter

.

If the zero is accepted the display will show ---- Tap the magnet over the

menu

switch to exit.

If the zero calibration was not within acceptable range the display will show F2. A recalibration is required; you
may need a cylinder of 20.9 air.

Note:

1. Carbon Dioxide sensors, except for high % range, should be zeroed in air due to the background levels of CO2

present. These sensors are best zeroed while being exposed to 100% nitrogen.
2. Where a purging gas has to be applied, use a flow rate of between 0.5 and 1.0 L/min. Allow sufficient time for
the sensor to respond.

5.1.2 Sensor Span

Tap the magnet over the

menu

switch, the display will change to Pass, Tap the magnet again on

menu

display will

show E 1.

Tap the magnet over the

up

switch, the display will change to E 2.

Tap the magnet over the

enter

switch the display will flash E: 2 and

XX (XX

= the calibration gas level).

Caution:

Be sure that the calibration gas you use matches the type and concentration of calibration gas display on the

EX-

6120, 6165 S

ERIES

IR

If the calibration gas you have does not match the displayed value, place the magnet over the

up

or

down

switch to

adjust the value.

Tap the magnet over the

enter

switch, the display will flash E: 2

Apply the calibration gas. The display will flash E: 2 and a number (sensor response to gas) when the number stops
changing,

Place the magnet over the

enter

switch. The display will show ---- if the calibration was successful.

Tap the

menu

switch to exit and remove the calibration gas.

Table 2: EX-6120, 6165 S

ERIES

IR Relative response curves based on a Methane Calibration

See Accessories Section 10 for

ENMET

Gas Cylinder part numbers.

1 2 3 4

MENU UP DOWN ENTER