0 calibration – ENMET EX-6165 User Manual
Page 12

EX-6120, 6165 SERIES IR
ENMET Corporation
9
5.0 Calibration
5.1 Sensor Calibration
This feature allows the sensor to be calibrated. Ensure that the correct sensor type is selected in the configuration prior to
calibration. Refer to section for details of the menu system operation.
5.1.1 Sensor Zero
Tap the magnet over the
menu
switch, the display will change to Pass, Tap the magnet again on
menu
display will
show E 1.
Tap the magnet over the
enter
switch the display will then alternate E: 01 Wait approximately 1 minute and Tap
enter
.
If the zero is accepted the display will show ---- Tap the magnet over the
menu
switch to exit.
If the zero calibration was not within acceptable range the display will show F2. A recalibration is required; you
may need a cylinder of 20.9 air.
Note:
1. Carbon Dioxide sensors, except for high % range, should be zeroed in air due to the background levels of CO2
present. These sensors are best zeroed while being exposed to 100% nitrogen.
2. Where a purging gas has to be applied, use a flow rate of between 0.5 and 1.0 L/min. Allow sufficient time for
the sensor to respond.
5.1.2 Sensor Span
Tap the magnet over the
menu
switch, the display will change to Pass, Tap the magnet again on
menu
display will
show E 1.
Tap the magnet over the
up
switch, the display will change to E 2.
Tap the magnet over the
enter
switch the display will flash E: 2 and
XX (XX
= the calibration gas level).
Caution:
Be sure that the calibration gas you use matches the type and concentration of calibration gas display on the
EX-
6120, 6165 S
ERIES
IR
If the calibration gas you have does not match the displayed value, place the magnet over the
up
or
down
switch to
adjust the value.
Tap the magnet over the
enter
switch, the display will flash E: 2
Apply the calibration gas. The display will flash E: 2 and a number (sensor response to gas) when the number stops
changing,
Place the magnet over the
enter
switch. The display will show ---- if the calibration was successful.
Tap the
menu
switch to exit and remove the calibration gas.
Table 2: EX-6120, 6165 S
ERIES
IR Relative response curves based on a Methane Calibration
See Accessories Section 10 for
ENMET
Gas Cylinder part numbers.
1 2 3 4
MENU UP DOWN ENTER