Wedge Quiet Interval

Product Reference Guide

73

Wedge Quiet Interval

This option specifies the amount of time to look for keyboard activity before the read-
er breaks the keyboard connection in order to transmit data to host. The selectable
range for this feature is from 0 to 990ms in 10ms increments.

Follow these instructions to set this feature:

1. Determine the desired setting in milliseconds.

2. Divide the desired setting by 10 (setting is in 10ms increments). Pad the result with lead-

ing zeroes to yield two digits. For example: 0 = 00, 5 = 05, 20 = 20, etc.

3. Scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Programming

Mode.

4. Scan the barcode: SELECT WEDGE QUIET INTERVAL SETTING.

5. Scan the appropriate two digits from the keypad in

Appendix E, Keypad

, that represent

the duration which was determined in the steps above. You will hear a two-beep indica-
tion after the last character.

6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit Programming

Mode.

This completes the procedure to set the Wedge Quiet Interval. See

Table 14

for some

examples of how to set this feature.

Table 14. Timeout Setting Examples

NOTE

This feature applies ONLY to the Keyboard Wedge interface.

NOTE

If you make a mistake before the last character, scan the CANCEL barcode to
abort and not save the entry string. You can then start again at the beginning.

STEP

ACTION

EXAMPLES

1

Desired Setting

10ms

150ms

600ms

850ms

2

Divide by 10 (and pad with leading
zeroes)

01

15

60

85

3

Scan ENTER/EXIT PROGRAMMING MODE

4

Scan SELECT WEDGE QUIET INTERVAL SETTING

5

Scan Two Characters From

Appendix E, Keypad

‘0’ and ‘1’

‘1’ and ‘5’

‘6’ and ‘0’

‘8’ and ‘5’

6

Scan ENTER/EXIT PROGRAMMING MODE