beautypg.com

Figure 3-13, 28 operations – ADLINK PXIe-9852 User Manual

Page 38

background image

28

Operations

in width, therefore the lowest sampling rate is 3.051KS/s (200MS/s
/ 65535).

Figure 3-13: Varying Sampling Rates by Adjusting Scan Interval Counter

Counter
Name

Length

Valid Value

Description

ScanIntrv

16-bit

1-65535

Timebase divider to achieve

equivalent sampling rate of the
digitizer, where Sampling rate =
Timebase / ScanIntrv

DataCnt

28-bit

1-268435452

Specifies the amount of data to
be acquired:

X

1 - 268435452 for pre-
trig or mid-trig mode
operation

X

1 - 268435452 for Data

Average mode for 1

channel

X

1 - 134217724 for Data

Average mode for 2

channel

trigDelayTicks 16-bit

1 -65535

Indicates time between a trigger
event and commencement of
acquisition. The unit of a delay
count is the period of the
Timebase.

TIMEBASE

DATA

D1

Acquisition
In Progress

Trigger

Acquisition is initiated following this clock edge

D1

ScanIntrv = 1

D2

D3

D4

D5

D6

D7

D8

D9

D10

D2

D3

D4

D5

D6

D1

D2

D3

D4

ScanIntrv = 2

ScanIntrv = 3