beautypg.com

2 calibration – ADLINK cPCI-9116 User Manual

Page 93

background image

Software Utility & Calibration

• 83

****** cPCI-9116 Utility Rev. 1.0 ******

Copyright © 2001-2002, ADLINK Technology Inc. All rights reserved.

: Calibration.

: Function testing.

: Quit.

>>> Select function key F1 ~ F2, or press to quit. <<<

Figure 22: CPCI-9116 Utility Main Screen

7.2 Calibration

In data acquisition processes, calibration of the measurement devices is very
important to maintain its accuracy. Users can calibrate the analog input
channels under the operating environment to optimizing the accuracy of the
9116 series card. The following section will guide you though the calibration
process for the 9116 series card.

Note: For an environment with frequently large fluctuations in temperature

and vibration, a 3 months re-calibration interval is recommended. For
laboratory conditions, 6 months to 1 year is acceptable.

7.2.1

What do you need

Before calibrating your 9116 series card, you should prepare the following
equipment’s:

A 5 1/2 digital multimeter (6 1/2 is recommended)

A voltage calibrator or a very stable and noise free DC

voltage generator