VACUUBRAND BVC professional G User Manual
Page 40

Adjustment routine for BVC professional with 4l polypropylene bot-
tle from software version V1.6, see rating plate
Action
Reaction
Meaning
Step 1
Touch and hold “4l P” key.
Case 1:
The LED of the “4l PP“
bottle symbol flashes
blue.
Adjustment is possible by
continuing to hold the “4l
PP” key, and proceeding
to step 2.
Case 2 / false alarm:
The alert tone for the false
alarm is shut off.
The LED of the “4l PP“
bottle symbol flashes
blue.
Step 2
When the LED of the “4l
PP” key flashes blue,
touch and hold the “bottle
change” key at the same
time.
A one-time buzzer
sounds.
After the buzzer, the LED
of the “bottle change” key
lights red.
Adjustment has been car-
ried out correctly.
Step 3
Release both “4l PP” key
and “bottle change” key. The LED of the “4l PP”
key lights blue.
The red LED of the “bottle
change” key turns off.
Adjustment finished.
Attention: Do not interrupt the adjustment routine once you have begun,
even if there is an alarm. Complete the entire routine, and repeat if neces-
sary.