With a new bottle, or to correct false alarms – VACUUBRAND BVC professional G User Manual
Page 38

At the factory, the level sensor is adjusted either to the 4l polypropylene
bottle or to the coated 2l glass bottle (G version) at standard conditions.
It may be necessary to adjust the sensor to compensate for changes in
bottle material, bottle size or changed ambient conditions.
Adjustment of the BVC level sensor to
work properly with a new bottle, or to
correct false alarms
NOTICE
Preparation
NOTICE
NOTICE
Use only the original VACUUBRAND BVC bottle or the original spare bot-
tle (see “Accessories - Spare parts”).
1. Do not switch off the BVC.
2. To shut off a false alarm prior to an adjustment, simply touch the
key marked “bottle change.” The red LED will continue to flash.
3. Empty the bottle.
4. Set the correct bottle size at the touch panel.
5. Wait approx. 1 minute before adjustment.
We also recommend emptying the bottle and adjusting the BVC profes-
sional if there is a false alarm indicating “liquid level” even though the
bottle is less than 3/4 full.
➨
If there is a false alarm in spite of an adjustment, repeat the adjust-
ment.
➨
If there is a false alarm in spite of
repeated adjustment, please call
+49 9342 808-5500 (after sales service), or your local VACUBRAND
sales office.
Pertains only to the BVC professional / BVC professional G