Filter and collection bottle, Filter, Notice – VACUUBRAND BVC professional G User Manual
Page 31

Filter
Midi
sart® 2000
0.20
µm PTFE
IN
sartorius stedim
The hydrophobic filter retains water and aqueous solu-
tions also in the form of aerosols. When using solvents or
due to evaporation of water, formation of condensate at
the outlet of the BVC is possible.
+
Important: Comply with applicable regulations when
disposing of solvents.
Filter and collection bottle
+
Attention: Do not remove connections in case of liq-
uid in the tube. Do not touch contaminated parts when
removing the screw cap. Risk of infection! Wear appro-
priate protective clothing, e.g., gloves.
Change filter if necessary in case of contami-
nation, weak aspiration or no aspiration.
+
Use only the original spare part (see section
“Accessories - Spare parts“) and sterilize
and dispose of the filter according to appro-
priate regulations.
Note: After disinfection with alcohol allow the
filter to dry completely.
NOTICE