During operation, Notice – VACUUBRAND BVC professional G User Manual
Page 28

The BVC professional (with 4l polypropylene
bottle) is equipped with quick couplings at the
bottle head (1) and at the inlet of the pump (2).
Quick couplings for other versions are avail-
able as accessories. (See section “Accessories
- Spare parts“.)
Quick coupling set (1): VHC
pro
to bottle:
When disconnected, the collection bottle clos-
es vacuum tight.
Quick coupling set (2): Bottle to pump unit:
When disconnected, the collection bottle closes
vacuum tight.
During operation
➨
Attention: When removing the aspiration
controller VHC
pro
liquid in the tube may leak!
Adopt suitable measures to prevent the re-
lease of dangerous, toxic, explosive, corro-
sive, noxious or polluting fluids, vapors and
gases. In such cases, use an appropriate
collecting and disposal system and take pro-
tective action for personnel, equipment and
environment.
+
Max. ambient temperature: 40 °C
• Use the system only with the
integrated hy-
drophobic filter to protect the vacuum sup-
ply from aspirated liquids and aerosols, and
to protect the environment/user from con-
tamination risk.
✔
1
2
max.
40°C
NOTICE