Cleaning and decontaminating, Notice, 121°c 2 bar absolute – VACUUBRAND BVC professional G User Manual
Page 35: Max. 80

Autoclaving
The collection bottle with bottle head and screw
cap, the quick coupling and the filter are des-
ignated for steam sterilization at 121°C and 2
bar absolute (1 bar overpressure). Time of ex-
posure according to DIN 58946 t
e
= 20 minutes.
Prior to autoclaving loosen or remove the
bottle head from the bottle.
It is the user´s responsibility to ensure effective
autoclaving.
Cleaning and decontaminating
121°C
2 bar absolute
The number of autoclaving cycles can be
marked on the plastic disc (Memory Disc) of the
filter (max. 20 autoclaving cycles according to
the manufacturer instruction).
Autoclaving
Maximum liquid level in collection bottle:
80%. When working with low boiling liquids or
liquids that tend to foam, the maximum liquid
level may be less.
max.
80%
121°C
2 bar
absolute
max. 20 x
NOTICE
NOTICE
Midi
sart® 2000
0.20
µm PTFE
IN
sartorius stedim
NOTICE
UV disinfection is permitted but may lead to discoloration
of the plastic parts.
Attention: UV disinfection acts only at the surface.