Troubleshooting The PSA-200U

Camera Not Found Error Message

If your laptop displays the following error

message (See Fig. 17), you need to...

Fig. 17

1. Shut the software off.
2. Disconnect power to camera and USB

cables.

3. Re-connect all power and USB ca-

bles.

4. Re-start the software and try again.

If you still receive the error message please

contact USL at (805) 549-0161 for technical

support.

Luminance values are off

If the Luminance values are off, they will

be off by a factor of 10. Check the Camera

Range (Settings in the Program menu).

Check to see if the neutral density filter is

attached to the camera lens if you have a

unit with the 30 - 300 foot-Lamberts (100

- 1000 cd/m2) option. (See the Program sec-

tion for details on correct operation of the

Camera Range settings.)

Computer message: Alignment Error!

Manually adjust the camera angle and the

camera zoom switch so that the image ap-

pears centered in the camera’s viewfinder.

Press F3, or click on Camera View. The im-

age should appear centered in the computer

display. Adjust the zoom switch on the cam-

era or the W / T with the mouse in “Camera

View” so the edges of the image are visible

in the computer display. Now press F2, or

click on Auto Align. If the error message in

Fig. 18 appears again, try using the Manual

Alignment function. If you cannot fix the

problem, call us for technical support.

Fig. 18

Computer message: Weave and Jitter

Warning - Screen luminance may be too

high/low for accurate measurement

Verify that you have at least six to 18 foot-

Lamberts of light from the lamp house.

If you cannot fix the problem, call us for

technical support.

Weave and Jitter measurements are not

constant

The initial period that the PSA takes to

sample the projected image is 30 seconds.

When this occurs, the Weave and Jitter read-

ings will be high or zero for a few seconds

until they are updated. This is an errone-

ous reading which is normal and should

be ignored. The Weave and Jitter function

should be left to run for two minutes in or-

der to obtain enough samples of the frame

to recognize samples of the film splice. The

film splice samples will then be discarded

automatically and will not be included in

the analysis. If you are having problems

that do not seem to be due to this, call us

for technical support.

Weave and Jitter measurements do not

seem to be accurate

Make sure the camera is zoomed in on the

small squares of the RP40 test pattern (see

Fig. 6). The camera should be aimed at the

center of the test pattern and the Analysis

Square (in the bottom left corner) should

be trained only on the small black and

white squares of the RP40 test pattern. Be

sure that the projector is properly focused

on the image. The pattern in the center of

the screen, generated by the Weave and

Jitter test film, cannot be used for analysis.