Error codes – Triner Scale TS-700 Series V2.0 User Manual

Page 36

background image

Operating Guide | TS-700 Series


A-15

Appendix For Scale Technicians

Error Codes

ERROR CAUSE

SOLUTION

uuuuuu

1. Overloaded

condition.

2. Wrong

connection at
load cell.

3. Load cell has

quality problem.

1. Reduce the weight.
2. Check load cell

connection

3. Inspect load cell. Check

the input and output.

nnnnnn

1. Calibration is

incorrect.

2. Bad connections.
3. Load cell has

quality problem

.

1. Check scale is resisted

or not, foot is kept level
or not.

2. Check load cell

connection.

3. Check input and output

resistance of load cell

.

ERR1

Overweight
calibration error.
Incorrect weight
input or incorrect
weight on
weighing
platform.

1. Input weight correctly

during calibration.

2. Place correct weight on

weighing platform.

ERR2

Underweight
calibration error.
Incorrect weight
input or incorrect
weight on
weighing
platform.

The calibration weights
Minimum is 10% of Max.
cap. Recommend 60%-
80% of Max. Cap.

ERR3

During
calibration, the
input signal is
negative.

Check all connections.
Check load cell.
If connections and load
cell are okay, PCB
needs replacing.