Operating Guide | TS-700 Series


A-2

Appendix For Scale Technicians

Calibration Switch, TS-700 MS: Remove the
screws and the back plate of the enclosure. A
white-capped calibration switch is located at the
bottom left corner of the circuit board. Press the
switch fully in and release. This will place the switch
in the up position, allowing access to the calibration
parameters. Loosely reattach the rear plate.

Calibration Switch, TS-700 SS: Remove the
screws and the back plate of the enclosure. A
calibration switch is located at the bottom left corner
of the circuit board. Move the switch to the “On”
position, allowing access to the calibration
parameters. Loosely reattach the rear plate.

Calibration Switch, TS-700 WB: Disconnect all
cables and cords. To remove the back plate of the
enclosure, place the indicator upside down (not face
down) on a work surface. Locate the two slotted
holes near the connection sockets. Insert a flat
blade screwdriver into each slot (one at a time),
pushing the blade inward while angled towards the
back of the indicator. This will release the retaining
prong and enable removal of the back plate.
Carefully lift the rear plate.

A white-capped calibration switch is located at the
bottom left corner of the circuit board. Press the
switch fully in and release. This will place the switch
in the up position, allowing access to the calibration
parameters. Loosely reattach the rear plate.