Operating guide | ts-700 series, Appendix for scale technicians – Triner Scale TS-700 Series V2.0 User Manual
Page 28

Operating Guide | TS-700 Series
A-7
Appendix For Scale Technicians
PARAMETER FUNCTION
OPTIONS
C06
Calibration
0 = skip calibration
1 = proceed with calibration
Description:
With C06 flashing, press the
key, then use the up
arrow key to change the
parameter setting to 1.
Press the
key to
proceed. “Span” will briefly
display, then all digits will
display.
Load the weighing platform,
then adjust the displayed
digits to match the weight on
the platform, i.e, 4,000.
Press the
key to
proceed. The indicator will
calibrate, then display
“CalEnd”.
PRESS THE
KEY TO
ACCEPT THE
CALIBRATION AND EXIT
SETTINGS MODE.
CAUTION -- IMPORTANT NOTICE!!
After calibrating, be sure to exit settings mode
without activating parameter C07!
Parameter C07 is the parameter that can DELETE all
custom settings and reset the indicator to factory
settings.