Operating Guide | TS-700 Series


A-7

Appendix For Scale Technicians

PARAMETER FUNCTION

OPTIONS

C06

Calibration

0 = skip calibration

1 = proceed with calibration

Description:

With C06 flashing, press the

key, then use the up

arrow key to change the
parameter setting to 1.

Press the

key to

proceed. “Span” will briefly
display, then all digits will
display.

Load the weighing platform,

then adjust the displayed
digits to match the weight on
the platform, i.e, 4,000.

Press the

key to

proceed. The indicator will
calibrate, then display
“CalEnd”.


PRESS THE

KEY TO

ACCEPT THE
CALIBRATION AND EXIT
SETTINGS MODE.

CAUTION -- IMPORTANT NOTICE!!

After calibrating, be sure to exit settings mode

without activating parameter C07!


Parameter C07 is the parameter that can DELETE all

custom settings and reset the indicator to factory
settings.