Calibration procedure – Triner Scale TS-700 Series V2.0 User Manual
Page 24

Operating Guide | TS-700 Series
A-3
Appendix For Scale Technicians
ALL MODELS: After enabling access to calibration
parameters, proceed as follows:
CALIBRATION PROCEDURE
1) Power on the indicator.
2) Press the PRINT and HOLD keys. “C” will
display. Release the keys.
3) “C01” will display, with the 1 flashing. The
indicator is now is Parameter Settings mode.
Navigating in Settings Mode
When the indicator is in settings mode, refer to the
graphic printed below the key to determine the
keypad key’s function.
The and arrow keys are used to
increase/decrease the value of the flashing digit, or
to navigate up and down a list of preset options.
The and arrow keys are used to move the
active digit over to the left or to the right.
The
key and
key can be used to move
forward or backward through the parameters. Also,
the
key is used to enter/accept changes after
they have been made to the parameter settings.
The
key can be used to exit settings mode.
With the indicator in settings mode and the
calibration weight nearby, proceed with the
calibration process while referring to the following
chart:
NOTE: After completing and verifying the
calibration, set the calibration switch to the “Off”
position and securely fasten the rear plate.