Operating Guide | TS-700 Series


A-1

Appendix For Scale Technicians

Calibrating a TS-700 Series Indicator

NOTICE: The TS-700 SS/MS/WB are NTEP
Class III Legal for Trade approved and
certified. Unlocking the calibration
parameters is not allowed by persons other
than qualified scale technicians.






ALL MODELS: Place the Indicator in Calibration
Mode and Determine if the Calibration
Parameters are Accessible
1) Power on the indicator.

2) Press the PRINT and HOLD keys. “C” will
display. Release the keys.

3) If the calibration parameters are accessible,
“C01” will display, with the 1 flashing. Proceed to
page 8, “CALIBRATION PROCEDURE”.

NOTE: If C08 Displays instead of C01, the
calibration switch must be adjusted. Disconnect the
AC adapter and power off the indicator.

DEFAULT SETTINGS CAUTION
Placing the indicator in calibration mode enables
access to the Default Settings Parameter, CO7.
NEVER activate Parameter C07 unless you are a
qualified scale technician,
and:
A) You understand that all custom settings will be
deleted permanently, and
B) You understand how to restore all custom
parameters to the required settings based on the
scale’s specifications and usage.