Warning, Danger, Important safety instructions – Oakworks CFLU401 User Manual

Page 6

background image

WaRnInG

To reduce the risk of burns, fire, electric shock or injury to persons:
1. Unplug this furnishing from the electrical outlet before cleaning.
2. Unplug from outlet before adding or removing parts.
3. Never operate this furnishing if it has a damaged cord or plug, if it is not working properly, if it has been dropped or

damaged, or dropped into water. Contact Oakworks Customer Service before use.

4. Keep the cord away from heated surfaces.
5. Never drop or insert any object into any opening.
6. Do not use outdoors.
7. Do not operate where aerosol (spray) products are being used or where oxygen is being administered.

DanGeR

The potential of exposure to harmful x-rays exists when this table is in use. The use of adequate x-ray barrier

devices is necessary to provide protection to both the operator and the patient. X-ray barrier devices are recommended

for the patient outside of the intended target area to prevent exposure to scattered radiation from the x-ray generat-

ing source.
The tabletop has a typical aluminum filtration equivalence of 1.10 mm as measured at 100 kVp and a half value (HVL)

of 3.6 mm. (NOTE: Bariatric Top has a typical aluminum filtration equivalence of 1.50 mm. as measured at 100 kVp

and a half value (HVL) of 3.6 mm.)

• O

akworks

®

Imaging Tables may be used for x-ray imaging where the x-ray generator is located below the

tabletop and the image receptor is located above the tabletop. This is the recommended method. The tables may

also be used for x-ray imaging where the x-ray generator is located above the tabletop and the image receptor is

located below the tabletop. This application will result in greater patient exposure to x-rays. The operator must

weigh this issue with the imaging requirements and patient exposure issues.

• The Imaging Tables may be used with the x-ray generator above the tabletop and a film cassette located on

the tabletop. (See contradictions regarding film cassette size.)

• The x-ray generator should never be located above the tabletop when the O

akworks

®

Imaging Table and

the O

akworks

®

Spine Positioning System/Prone Pillow are used together. This type of use requires that the x-ray

generator is located below the tabletop and the imaging intensifier or film cassette located above the tabletop.

3

IMPoRTanT saFeTy InsTRUCTIons