Specifications – Oakworks CFLU401 User Manual

Page 38

background image

35

sPeCIFICaTIons

Guidance and Manufacturer’s Declaration - Immunity all equipment and systems

The table is intended for use in the electromagnetic environment specified below. The customer or the

user of the table should assure that it is used in such an environment.

Immunity Test

IeC 60601 test level

Compliance level

electromagnetic

environment - guidance

electrostatic discharge

(esD)

eN/IeC 61000-4-2

±6 kv contact

±8 kv air

±6 kv contact

±8 kv air

Floors should be wood,

concrete or ceramic tile.

If floors are covered with

synthetic material, the rela-

tive humidity should be at

least 30 %.

electrical fast transient/

burst

eN/IeC 61000-4-4

±2 kv for power supply lines

±1 kv for input/output lines

±2 kv for power supply lines

Not applicable for input/

output lines

Mains power quality should

be that of a typical com-

mercial or hospital environ-

ment.

surge

eN/IeC 61000-4-5

±1 kv line(s) to line(s)

±2 kv line(s) to earth

±1 kv line(s) to line(s)

±2 kv line(s) to earth

Mains power quality should

be that of a typical com-

mercial or hospital environ-

ment.

voltage Dips/Dropout

eN/IeC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles
<5 % U

T

(>95 % dip in U

T

)

for 5 sec

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

Mains power quality should

be that of a typical com-

mercial or hospital environ-

ment.

If the user of the table re-

quires continued operation

during power mains inter-

ruptions, it is recommended

that the table be powered

from an uninterruptible

power supply or a battery.

Power frequency (50/60

Hz)

Magnetic Field

eN/IeC 61000-4-8

3 A / m

3 A / m

Power frequency magnetic

fields should be at levels

characteristic of a typical

location in a typical com-

mercial or hospital environ-

ment.

NoTe U

T

is the a.c. mains voltage prior to application of the test level.