beautypg.com

Specifications product specifications, Environmental conditions, Electrical specifications – Oakworks CFLU401 User Manual

Page 36

background image

33

sPeCIFICaTIons

PRoDUCT sPeCIFICaTIons

NorTH AMerICA

eUroPe

Weight

359-452 lbs.

163-205 kg.

shipping Weight

452-544 lbs.

205-247 kg.

Lifting Capacity

500 lbs. (750 lbs. Bariatric)

227 kg. (340 kg. Bariatric)

Materials of Construction

This product contains no latex.

Aluminum Filtration

equivalence and thick-

ness of Carbon Fiber

Tabletop

ALUMINUM eqUIvALeNCe

THICkNess

Non-Bariatric Tops

100 kvp, HvL of 3.6mm = 1.10 (1.35 max)

1.65” (4.2 cm)

Bariatric Tops

100 kvp, HvL of 3.6mm = 1.50 (1.79 max)

2.18” (5.5 cm)

All extensions

100 kvp, HvL of 3.6mm = 1.66 (2.00 max)

.5” (1.3 cm)

1” (2.5 cm.) Table Pad

100 kvp, HvL of 3.6mm = .88 (1.07 max)

1” (2.5 cm)

2” (5 cm.) Table Pad

100 kvp, HvL of 3.6mm = .88 (1.07 max)

2” (5 cm)

This product complies with United states Department of Health and Human services radiation performance

standards, 21 CFr sub chapter J, in effect at the time of manufacture for radiographic tables.

enVIRonMenTaL ConDITIons

Conditions

Temperature

Humidity

Atmospheric Pressure

Normal Use

50° (10°C) to 104° (40°C)

20% to 60% rH

98 to 105 kPa

storage & Transport

-20° (-29°C) to 135° (57°C)

20% to 95% rH

98 to 105 kPa

eLeCTRICaL sPeCIFICaTIons

CFPM100 Tables only

CFPM400, CFPM401, CFPM301,

CFLU401, CFUr401 & CFUr301

Tables

Designed for:

NorTH AMerICA

eUroPe

NorTH AMerICA

eUroPe

Input service (All electric

tables)

120 vAC/15

amp/60 Hz

230 vAC/10

amp/50/60 Hz

120 vAC/15

amp/60 Hz

220 vAC/10

amp/50/60 Hz

Current Draw

5.8 amps

3.0 amps

8.2 amps

4.5 amps

Maximum Momentary

Current Consumption

9.0 amps

4.5 amps

12.3 amps

6.7 amps

voltage output to Ac-

tuators

120 vDC

220 vDC

24v top,

120v Column

24v top,

220v Column

electric shock Protection

Class 1 equipment

Tabletop Applied Part

Type B Applied Part

Ingress Protection rating

IPX0

Mode of operation

Intermittent operation: MAX 2 minutes oN / 18 minutes off

This manual is related to the following products: