Instruction manual fh, Sauter gmbh – KERN & SOHN FH User Manual
Page 2
Sauter GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: [email protected]
Tel: +49-[0]7433- 9933-199
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Instruction Manual
FH
FH_SI-BA-e-1317
2
4.
Technical Data
-
Accuracy: ± 0,5 % of Capacity (measuring range)
-
Data Sampling Rate: 2.000 Hz
-
Weight: 640 g
5. Operation
Display
(1) Measuring Result
(2) Measuring Units
(3) Activation of PRINT Function
(4) Indication of power charging status
(5) PEAK or AUTO-PEAK Mode
(6) Average value of stored peak values
(7) Force direction
(8) Occupancy of storing spaces
(9) AVERAGE- or Saving Mode
Operation keys
ON / OFF:
ON / OFF key
(For ON, press 1 sec.)
UNIT:
-
Press shortly:
Select unit:
N, kg or lb
-
Press for 2 sec.:
Display return
ZERO:
Three functions:
-
Zeros the measuring result (Tara function)
-
Cleans the peak value (in Peak mode)
-
Saves a setting (in SET mode)
SET:
-
1 x Press: Upper Limit [HidT / Hi Lt]. To
change press: ▲ or ▼ (see section
High/Low Limit function)
-
1 x Press: Lower Limit [LodT]. To change
press ▲ or ▼ (see section High/Low Limit
function)
-
1 x Press: Minimum limit to save peak
values in the instrument’s memory.
[LE.SET / Lo Pe].
Please see section „Min
Limit Peak Save“. To change press: ▲ or
▼ (Only active in „Peak-Mode“) (see
section min. limit peak save)
-
1 x Press: Sets the Auto-STOP force value
that stops the movement of a test stand.
Adjustment of the force value. To change
press: ▲ or ▼ (only active in „Battery-
Mode“)
1 x Press: Auto-OFF Function. Turns the
instrument off after a here defined time
period in sec.
[P.OFF]. To change press: ▲
or ▼ (Only active in „Battery-Mode“)
-
1 x Press: Peak-Freeze-Time [PE.2E /
A.PE / HoldT]: Time period (in sec.) in
which a peak value is being shown in the
display in sec.
To change press: ▲ or ▼
-
1 x Press: Data Transfer [rS232] to PC
(PC) or to the printer (Print) or (in version U
5.1) to the test stand
- PC: Pressing SET once: Saves the Settings
- PRINT: Pressing SET twice: sending data to the printer
- STAND: Pressing SET three times: sending a signal to
the Test Stand to stop its movement (in version U 5.1)
BACK LIGHT:
Hi Lt
Lo dT
Lo Pe
P.OFF
HoldT
rS232
StoP