19.1 Error messages
Error message
Description
Possible reasons/
solution
Err 4
Zeroing range exceeded
due to switching-on balance
or pressing
(normally
4% max)
• Object on the weighing
plate
• Overload when zeroing
• Inappropriate
adjustment
• Damaged weighing cell
• Damaged electronics
Err 5
Keyboard error
• Unpurposeful
operation of the scale
Err 6
Value outside the A/D
changer range
• Weighing plate not
installed
• Damaged weighing cell
• Damaged electronics
Err 19
Zero point displaced
• Remedy:
Adjust / linearize
FAIL H / FAIL L
Adjustment error
• Inappropriate
adjustment
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.
84