(1) Adjusting the bulk scales, all models except for CFS 50K-3
1. Internal resolution
Press
, the internal resolution will be shown.
Return to menu by
.
Use
to select the next menu item
„Cap“
.
„Cnt“
2. Position decimal point / capacity
„CAP“
Press
on display „CAP“; the currently adjusted position
of the decimal point will be displayed.
Select desired setting with
and acknowledge by
.
The currently set capacity is displayed.
Delete with
for changes, and input desired value with
the numeric keys.
Confirm input with
, the balance returns to the menu.
Select the next menu point „div“ with
.
„dESC“ „0.00“
„SEL“ „000030“
„CAP“
3. Readability
„div“
Press
, the current setting will be displayed.
Select desired setting by
and confirm with
, the
balance returns to the menu.
Select the next menu point „AZt“ with
„inC“ „1“
„div“
4. Automatic zero tracking
on display change
„AZt“
Press
, the current setting will be displayed.
Select desired setting by
and confirm with
, the
balance returns to the menu.
Select the next menu point „0 AUto“ with
.
„AZn“ „2d“
„AZt“
CFS/CCS-BA-e-1422