13 configuration of bulk scale, Configuration of bulk scale – KERN & SOHN CFS 50K-3 User Manual
Page 55

13 Configuration of bulk scale
Changes may only be carried out by trained specialized personnel.
Ex factory, the scales KERN CFS respectively counting systems KERN CCS have
been programmed such that no changes are generally necessary.
But if there are special conditions of use or if as bulk material balance an other
weighing bridge (not preconfigured by KERN) is connected, in the menu block
"
" the required settings can be made.
Technical data
Supply voltage:
5 VDC
Max. signal voltage
0-20 mV
Zeroing range
0-5 mv
Sensitivity
> 0.02
μv
Resistance parameter
87 O Min., 4 x 350 Ω load cell
Connection
4 poles
Max. cable length
6 meter
Connection plug
9 pin d-subminiature bushing
Navigation in the menu:
With
or
(model CFS 50K-3), the individual menu items may be
selected one by one.
Confirm selected menu item with
or
(model CFS 50K-3). The current
setting will be displayed.
With
bzw.
(model CFS 50K-3) switch over to the available settings.
Either save with
bzw.
(model CFS 50K-3) or reject with
or
(model CFS 50K-3).
CFS/CCS-BA-e-1422
55