(2) Adjustment of bulk scales, model CFS 50K-3
1. Internal resolution
Press
, the internal resolution will be shown.
Return to menu by
.
Select the next menu point „dESC“ with
.
„Cnt“
2. Position decimal point /
„dESC“
Press
on display „
dESC
“; the currently adjusted position
of the decimal point will be shown.
Select desired setting with
and acknowledge by
.
Use
to select the next menu item [CAP].
„dESC“ „0.00“
CAP
3. Capacity
Press
on display „CAP“; the currently adjusted capacity
will be shown.
Select desired setting with
and acknowledge by
.
Delete with
for changes, and input desired value with
the numeric keys.
Confirm input with
, the balance returns to the menu.
Select the next menu point „div“ with
.
„CAP“
„SEL“ „060.000“
„CAP“
4. Readability
„div“
Press
, the current setting will be displayed.
Select desired setting by
and confirm with
, the
balance returns to the menu.
Select the next menu point „AZt“ with
„inC“ „5“
„div“