15.3 Remote control instructions
Attention
:
Inputting characters and command codes not shown here into the balance may not
only alter the previous settings but may also impair proper measurement.
If by mistake characters or commands not shown here are entered into the balance,
immediately unplug the power supply cable and wait about ten seconds before plug-
ging again.
Command Code
Function
Description
D01
Continuous output
The balance continuously outputs every 110ms .
D05
1 time output
Corresponds to [PRINT] key
D06
Auto Print
See 14.4
D07
1 time output with stability
information
The status of the stability mark is appended to the head
of the data with output.
S: when the standstill display appears
U: when standstill display does not appear
D08
1 time output at stability
After command input, the data are output at the first
appearance of the standstill display.
D09
Halt output
Auto Print and continuous output halted
Q
ON/OFF switching
Switches between standby status and measurement
status.
T
Taring
Corresponds to the [O/T] key
TS
Taring after stability wait
After command input, taring is done at the first appear-
ance of the stability mark.
C18
Measurement span calibration
+
Measurements in add-up mode
See ch. 14.5
R
Total reset
All application measurements terminated and reset
mg
mg unit
PERCENT
Percent determination
PCS
Parts counting
CT
ct unit
SDENCE
Density of solids
LDENCE
Density of liquids
%
100% setting
G
g, % switching
- g
g unit removal
- mg
mg unit removal
- PERCENT
Deleting percentage determina-
tion t
- PCS
Piece counting removal
- CT
ct unit removal
- SDENCE
Solid specific gravity removal
- LDENCE
Liquid specific gravity removal