14.5 Add up mode
An optional number of single weighing processes are automatically added up to a
total sum, e.g. all single weighing processes of a batch or if you have to carry out a
great number of measurements of small samples.
When the standstill control ( ) is complete the weighing value is automatically is-
sued to the optional printer. The displayed value is added into the total adding
memory. Afterwards automatic taring will take place. This is repeated newly every
subsequent time a new sample is placed on the weighing pan. When the last single
weighing process is finished, press the [ON/OFF] key to receive the total sum ("TO-
TAL=").
Important:
This mode can be applied to any unit.
The initial load of the balance must be kept within the ± 5-fold of the ze-
ro range.
Weight of sample ≥ 10 x zero range (zero range = 0.25 e)
If during the use of multiple range balances the smallest readability digit
was turned off by pressing the [1d/10d] key, evaluation is based on the
latest add-on value displayed before.
Activate function:
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press [TARE] key
(Example)
Repeatedly press the [CAL] key until the current "Addon:**" set-
ting appears.
Press
[TARE]
key
You can use the [CAL] key to select between the following set-
tings:
“
Adon-on”
Function activated
“ Adon-oF”
Function deactivated
The current setting is indicated by the standstill display ( ).