14.3 Density determination
Density determination with the help of the flush-mounted platform facility is described
in the following.
Density determination becomes even easier when an optional set for density deter-
mination is applied. For further information please refer to the operating instructions
enclosed with the set for density determination.
1. Remove the below-weigh hook cover on the floor of the balance, after remov-
ing the two fixing screws.
2. Hang the user-prepared hanging pan from the below-weigh hook and sink that
hanging pan into the sample liquid in the tank.
14.3.1 Density determination of solids
Condition
: Function ",d" (density of solids) is activated. See Chapter 11.3.
Press the [CAL] key repeatedly until "SettinG" appears.
Press the
[TARE]
key.
Repeatedly actuate the
[CAL]
key until "LSG SEt" appears
(Example)
Press the [TARE] key. The currently set density for the liquid to be
measured appears. In the upper part of the display panel, the
symbol and the # symbol appear in order to indicate numer-
ical input status. The leftmost digit blinks.
(Example)
Enter density for your liquid to be measured. When the [UNIT] key
is pressed, the numerical of the blinking digit increases by 1 at a
time. You can determine the value of the flashing digit, or shift the
flashing digit by one position to the right, by pressing the [PRINT]
key. Confirm your setting by pressing the [TARE] key.
Repeatedly press the [ON/OFF] key until the balance is in weigh-
ing mode.