KERN & SOHN ABT 220-5DM User Manual
Page 6

ABT-BA-e-1312
6
KERN
ABT 120-5DM
ABT 220-5DM
Readability (d)
0.01/0.1 mg
0.01/0.1 mg
Weighing range (max)
42 g/120 g
82 g/220 g
Minimum load (Min)
1 mg
1 mg
Verification value (e)
1 mg
1 mg
Verification class
I
I
Reproducibility
±0.02/ 0.1 mg
±0.05/ 0.1 mg
Linearity
± 0.05/0.2 mg
± 0.1/0.2 mg
Stabilization time
3sec./10 sec.
Adjustment weight
internal
Weighing units
(calibrated appliances)
g, ct
Smallest part weight for piece
counting
1 mg
Reference quantities at
piece counting
10, 20, 50,100
Weighing plate, stainless
steel
ø 80mm
Dimensions of the housing (B
x D x H)
217 x 356 x 338
Dimensions
Glass wind screen [mm]
Weighing room 168 x172 x223
Net weight (kg)
7
Permitted environmental
condition
+10° C to +30° C
Humidity of air
max. 80 % relative (not condensing)
AC adapter (Primary)
AC 100 -240 V, 400 mA 50/60Hz
Rated electric power supply
DC 12 V, 1 A
Pollution Degree
2
Overvoltage Category
Category II
Altitude
Up to 2000 m
Installation Site
device may only be used indoors