15 Data output

15.1 RS 232C interface
Pin allocation of balance output plug:

PIN No.

Signal

2

TXD

3

RXD

6

DSR

7

SG

20

DTR

5

CTS

4

RTS

15.2 Data Formats
Note:  space character and DL the delimiter code.

1. Input Data Format

COMMAND CODE + DL (see 15.2.3)

2. Output data format

In weighing mode:

(ex.)

S -200.000gDL

stability

polarity

Polarity data

Positive: space character (()

Negative: minus (-)

Stability information

(when output includes stability information)

Stable : S

Unstable : U

In “oL” or “-oL” display

(ex.)

U -oLDL

stability

polarity

Polarity data

Positive: space character ()

Negative: minus (-)

Stability data

(when output includes stability information)

Stable : S

Unstable : U

3. Data format

ASCII (JIS) code

Baud rate, parity (and bit length), delimiter, stop bit, format, and handshake

differ by menu item selections.